The variation of pH in perfusate during renal machine perfusion has been regarded exclusively as an effect of kidney metabolism. During preservation of 21 canine kidneys for 48 hours in a Gambro PF-2D machine the pH was measured in the human albumin perfusion medium used, and in control samples stored a 4 degrees C. An increase in pH from 7.21 (+/- 0.03) to 7.66 (+/- 0.10) was found in the machine perfusate. In the stored perfusion medium an increase from 7.21 (+/- 0.03) to 7.43 (+/- 0.20) was measured. We conclude that the pH change in perfusate is not a reliable parameter for the functional state of the kidney unless the spontaneous pH variation in the perfusion medium is simultaneously recorded and a correction made.
No takes yet. Share an insight, caveat, or question.
Olsen et al. (1980) studied this question.