Key result
MiCardia enCorSQ device enables late-stage adjustable annuloplasty for ischemic MR via temperature-induced diameter reduction.
Describes the surgical technique for implanting and activating the adjustable MiCardia enCorSQ annuloplasty device for ischemic mitral insufficiency.
May enable late adjustable annuloplasty in ischemic MR; leaves open durability and outcomes in prospective trials.
Presentation of the MiCardia enCorSQ annuloplasty device implantation technique for ischaemic mitral regurgitation including late-stage activation: exposure of mitral valve using video-assisted right lateral mini-thoracotomy, annuloplasty, tunnelling of permanently attached lead through the left atrial wall and subcutaneous implantation; and late activation: minor cut-down procedure for lead exposure, connection to proprietary energy source (MC-100 RF generator), echocardiography and fluoroscopy guidance, anterior-posterior diameter reduction of memory-shape alloy core annuloplasty device by raising the temperature a few degrees above body temperature.
No takes yet. Share an insight, caveat, or question.
Czesla et al. (2013) studied Ischaemic mitral insufficiency. MiCardia enCorSQ annuloplasty device was evaluated. The MiCardia enCorSQ annuloplasty device implantation technique enables late-stage adjustable mitral annuloplasty for ischaemic mitral regurgitation via temperature-induced diameter reduction.