Answers are grounded in this paper and cite their sources.
A sensitive high-performance liquid chromatographic method with fluorescence detection to determine 3,4-methylenedioxymethamphethamine (MDMA) and 3,4-methylenedioxyamphethamine (MDA) in human and rat whole blood or plasma samples was developed by using 4-(4,5-diphenyl-1H-imidazol-2-yl)benzoyl chloride (DIB-Cl) as a label. MDMA and MDA in a small amount of blood sample (ca 100 microL) were extracted by liquid-liquid extraction with ethyl acetate, and were derivatized with DIB-Cl under mild conditions (10 min at room temperature). A good separation of DIB-derivatives could be achieved within 45 min using a commercially available ODS column with an isocratic eluent of 10 mM citric acid-20 mM Na(2)HPO(4) aqueous buffer (pH 4.0)-CH(3)CN-CH(3)OH (50:45:5, v/v/v %). The calibration curves prepared with 1-methyl-3-phenylpropylamine (MPPA) as an internal standard showed good linearity (r = 0.999) with 0.36-0.83 ng/mL detection limit at a signal-to-noise ratio of 3. MDMA and MDA in rat whole blood could be monitored for 6 h after a single administration of MDMA (2.2 mg/kg, i.p.). The pharmacokinetic parameters for MDMA and MDA obtained by triplicate measurements were 426 +/- 23 and 39 +/- 6 ng/mL (C(max)), 20 +/- 5 and 100 +/- 10 min (T(max)), respectively.