AIMS: To evaluate the electronic nose (EN) as method for the identification of ten clinically important micro-organisms. METHODS AND RESULTS: A commercial EN system with a series of ten metal oxide sensors was used to characterize the headspace of the cultured organisms. The measurement procedure was optimized to obtain reproducible results. Artificial neural networks (ANNs) and a k-nearest neighbour (k-NN) algorithm in combination with a feature selection technique were used as pattern recognition tools. Hundred percent correct identification can be achieved by EN technology, provided that sufficient attention is paid to data handling. CONCLUSIONS: Even for a set containing a number of closely related species in addition to four unrelated organisms, an EN is capable of 100% correct identification. SIGNIFICANCE AND IMPACT OF THE STUDY: The time between isolation and identification of the sample can be dramatically reduced to 17 h.
No takes yet. Share an insight, caveat, or question.
Moens et al. (2006) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: