Answers are grounded in this paper and cite their sources.
An electrochemiluminescent (ECL) lead biosensor was developed based on GR-5 lead-dependent DNAzyme for lead recognition and intercalated ruthenium tris(1,10-phenanthroline) (Ru(phen)(3)(2+)) as the ECL probe. The thiol-modified substrate was first immobilized on the surface of the gold electrode via gold-sulfur self-assembly. Subsequently, the hybridization of DNAzyme and its substrate and the automatic intercalation of Ru(phen)(3)(2+) proceeded. Intercalated Ru(phen)(3)(2+) can transfer electrons through double-stranded DNA to the electrode and its electrochemiluminescence was excited with a potential step using tripropylamine as the coreactant. In the presence of lead, the substrate cleaves at the scissile ribo-adenine into two fragments. The dissociation of DNAzyme occurs, leading to the releasing of intercalated Ru(phen)(3)(2+) accompanied by a decrease in the intensity of electrochemiluminescence. A quantity of lead can be calculated from this decrease. The biosensor is highly sensitive and specific, along with an ultra-low limit of detection of 0.9 pM and a dynamic range from 2 to 1000 pM. It enables analysis of trace amounts of lead in serum samples. The combination of the intercalated-Ru(phen)(3)(2+) ECL probe and the cofactor-dependent DNAzyme may push the performance of cofactor-sensing tactics to the extreme.
No takes yet. Share an insight, caveat, or question.
Gao et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: