Key result
Applying a beat-to-beat quality indicator to carotid PWV improves pulse pressure correlation by ~86%.
Why the study?
Local carotid PWV estimation is challenged by complex anatomy, wave reflections, and BP- and position-dependent wave interference causing nonlinearity, with a lack of beat-to-beat quality assessment.
Observational (n=10)
A novel beat-to-beat quality indicator improves the reliability of local carotid pulse wave velocity estimation by accounting for nonlinearity.
Quality filtering may refine carotid PWV estimates; leaves open whether it enhances risk stratification in prospective cohorts.
Background Local carotid pulse wave velocity (PWV) estimation is challenging due to a complex anatomy and associated wave reflections [1]. Wave interference is both blood pressure (BP) and position dependent, hence causing nonlinearity in the spatiotemporal PWV estimation [2]. In lack of beat-to-beat quality assessment [3], we propose a method to quantify this nonlinearity and thereby introduce a quality indicator (QI) for reliable PWV estimation. Methodology Carotid artery ultrasound (Vantage64, VerasonicsInc., USA) was acquired from 10 subjects (38 ± 10 years) in six repeated measurements. Simultaneously, ECG (ECG100C, BIOPACInc., USA) and noninvasive BP (NOVA, FinapresMedicalSystems B.V., NL) were recorded. Signals were processed into distension waveforms to extract systolic foot (SF) and dicrotic notch (DN) fiducials [4]. Beat-to-beat PWV estimates at SF (PWVSF) and DN (PWVDN) were obtained via spatiotemporal fitting of respective fiducial timings against transducer element distances. The proposed nonlinearity QI was quantified as standard deviation of the error between full-length PWV estimate and piecewise PWV estimates from the proximal, middle and distal subset of transducer elements. Linear regression was performed between independent pulse pressure (PP, in 2 [mmHg] intervals) and dependent PWV (means of all estimates) to validate QI according to Bramwell-Hill [5]. Results When qualifying PWV estimates with QI ≤1 [m/s] we observed a significant R2 increase for both PWVSF (0.22 to 0.41, with decreased RMSE from 0.47 to 0.42 [m/s]) and PWVDN (0.21 to 0.53, although with constant RMSE of 0.32 [m/s]) as shown in annexed figure. Conclusion The proposed beat-to-beat QI accounts for significant variability in the PWV-PP relation, albeit at substantial data reduction. Therefore, the QI may facilitate real-time quality assessment towards reliable local PWV estimation. The unexplained variability remainder requires further investigations for confounding variables.
No takes yet. Share an insight, caveat, or question.
Beutel et al. (2019) reported an observational. Beat-to-beat quality indicator (QI ≤1 m/s) vs. Unqualified PWV estimates was evaluated on Correlation (R2) between pulse pressure and PWV at systolic foot (PWVSF). Applying a beat-to-beat quality indicator (QI ≤1 m/s) to carotid pulse wave velocity estimates significantly increased the correlation (R2) with pulse pressure from 0.22 to 0.41.