We develop rapid chemical-vapor sensors based on optofluidic ring resonators (OFRRs). The OFRR is a glass capillary whose circular wall supports the circulating waveguide modes (WGMs). The OFRR inner surface is coated with a vapor-sensitive polymer. The analyte and polymer interaction causes the polymer refractive index to change, which is detected as a WGM spectral shift. Owing to the excellent fluidics, the OFRR exhibits subsecond detection and recovery time with a flow rate of only 1 mL/min, a few orders of magnitude lower than that in the existing optical vapor sensors. The detection limit is estimated to be 5.6 x 10(-6) refractive index units, over ten times better than other ring-resonator vapor sensors. Ethanol and hexane vapors are used as a model system, and chemical differentiation is demonstrated with different polymer coatings.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2008) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: