The cardioactive drug lidocaine was measured by a homogeneous enzyme immunoassay procedure (EMIT ®) using a procedure similar to the one described for the determination of antiepileptic drugs in serum. The assay requires 50 µl of serum per determination and less than 5 minutes to perform. The assay is specific for lidocaine and does not cross-react with lidocaine metabolites or other cardioactive drugs. Twenty replicate analyses of a serum containing lidocaine gave a CV of 3.4%. Results attained with a large series of patients administered lidocaine alone or with other cardioactive drugs were compared with results by a gas liquid chromatographic procedure. The correlation coefficient (r) was 0.976. No false postives or negatives were encountered.
No takes yet. Share an insight, caveat, or question.
Clifford B. Walberg (1978) studied this question.