High Resolution Image Download MS PowerPoint Slide Cyclic ion mobility–mass spectrometry (cIM-MS) is a powerful technique for separating and identifying isomeric mixtures of compounds. When coupled with chromatography, cIM-MS creates a multidimensional separation system, with high resolving power and peak capacity. In this study, we report the cyclic ion mobility separation and high-resolution mass spectrometry identification of four regioisomers of a Sugammadex-related impurity, abbreviated as Di-OH-SGM. Separation using multipass cyclic ion mobility was achieved by selecting the [M + 2Na] 2+ ion, while other adducts, such as [M + Na] +, [M + 2H] 2+, [M + H + Na] 2+, and [M – 2H] 2– did not yield isomer separation. Two methods were developed for ion mobility separation of the isomers: a conventional multipass method and a slicing method. Isomer assignment was based on the characteristic fragment ions. The collision cross section values ( cTW CCS N2 ) of the resolved cyclodextrin isomers were also determined. Ion mobility separation of structurally different fragment ions was demonstrated. Additionally, by coupling cIM-MS with reversed-phase liquid chromatography (HPLC-cIM-MS), two-dimensional separation of the isomers was achieved. The isomers, separated using HPLC-cIM-MS, were identified with the same approach as with cIM-MS alone, and their elution order provided insights into their relative hydrophobicity.