Key result
Remote monitoring of pacemakers and ICDs generates frequent technical alerts but fewer clinical alerts.
Observational (n=282)
Yes
In remote monitoring of pacemakers and ICDs, centres tend to prioritize critical technical data over patients' clinical profiles, highlighting the need for personalized alert settings.
Answers are grounded in this paper and cite their sources.
Optimizing alert settings may reduce clinic workload in remote CIED monitoring; leaves open whether tailored thresholds improve outcomes.
AIMS: The remote monitoring of pacemakers and implantable cardioverter defibrillators (ICDs) promotes accurate supervision of the patient and device. Alert settings appear to be a crucial parameter of its efficacy. The purpose of our study was to evaluate various settings for alerts and alert message management in patients with pacemakers and ICDs. METHODS AND RESULTS: We prospectively evaluated 282 patients (181 males, 101 females, mean age 72 ± 11 years) who were referred to nine electrophysiological centres in the Veneto region of Italy for a pacemaker (76 patients) or ICD (206 patients) implant in which remote monitoring was activated (Biotronik Home Monitoring®, Biotronik, Berlin, Germany). During a mean follow-up of 382 ± 261 days, we received 106,039 transmissions. In the pacemaker group, the alerts that were most frequently activated were those concerning battery exhaustion, surveying impedances, sensing and threshold measurements, as well as missing transmissions. In the ICD/ICD-cardiac resynchronization therapy group, the alerts nearly always activated were those concerning a detection setoff, battery exhaustion, critical values of impedance, or ineffective maximum energy shock. In both groups, the alarms for heart rate monitoring and supraventricular arrhythmia were activated in fewer cases at higher variability among centres. CONCLUSIONS: Our study demonstrates that more attention is paid to critical technical data than to patients' clinical profiles, probably to limit an excessive flow of data into the centre. Accurate alert settings, personalized to the patients' features, are essential for easier and more effective management of patients who are followed remotely.
No takes yet. Share an insight, caveat, or question.
Folino et al. (2011) conducted an observational in Pacemaker or ICD implantation (n=282). Remote monitoring alert settings was evaluated on Alert settings and alert message management. Remote monitoring of 282 patients with pacemakers and ICDs generated 106,039 transmissions, revealing that technical alerts were frequently activated while clinical alerts were used less often.
Synapse has enriched 10 related papers. Consider them for comparative context: