Key result
Biodegradable PLLA sirolimus-eluting stents reduce neointimal area ~40% vs Cypher stents in a porcine model.
Why the study?
Do biodegradable polymer-based sirolimus-eluting stents reduce neointimal area compared to permanent polymer stents in a porcine model?
Do biodegradable polymer-based sirolimus-eluting stents reduce neointimal area compared to permanent polymer stents in a porcine model?
Biodegradable PLLA-based sirolimus-eluting stents demonstrated greater long-term efficacy in reducing neointimal area compared to permanent polymer stents in a porcine model.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating preclinical data on biodegradable SES formulations; leaves open clinical superiority over permanent polymer stents.
Biodegradable stent coatings were recently introduced as a potential solution to overcome sustained inflammatory responses observed with permanent polymer-based drug-eluting stents. In a preliminary study, selected biodegradable or permanent polymer-based sirolimus-eluting stent (SES) formulations were screened for effectiveness in comparison to bare metal stents (BMS) at 28 days. Subsequently, the most favourable SES formulation was compared to commercially available SES (Cypher™) at 28, 90 and 180 days to investigate the histopathologic response as well as tissue, blood and organ pharmacokinetics. Overlapping SES implantation was conducted to evaluate vascular healing at 28 days in this particular setting. SES with biodegradable poly (L-lactide) polymer (PLLA) or poly(lactide-co-glycolide) showed the most favourable outcome with regards to reductions in neointimal area in comparison to BMS at 28 days. The PLLA SES showed a similar reduction in neointimal area compared to Cypher™ at 28 days, with significant greater reductions at 90 and 180 days (1.7 ± 0.7 mm² vs. 3.1 ± 1.5 mm², p=0.03 and 1.8 ± 1.2 mm² vs. 3.0 ± 1.5 mm², p=0.01, respectively). Sirolimus vascular tissue concentrations were detectable up to 90 days following implantation. Overlapping stented segments showed favourable histopathologic results with respect to fibrin deposition and endothelialisation at 28 days. In conclusion, the use of PLLA as drug-eluting matrix resulted in mild inflammatory responses in the presence of effective sirolimus tissue concentrations. The greater efficacy observed at long-term follow-up in PLLA SES compared to Cypher™ may be a multifactorial result of stent design, polymer biocompatibility and improved release kinetics.
No takes yet. Share an insight, caveat, or question.
Koppara et al. (2012) studied Coronary stent implantation (porcine model). Biodegradable poly (L-lactide) polymer (PLLA) sirolimus-eluting stent vs. Cypher™ sirolimus-eluting stent was evaluated on Neointimal area at 180 days (p=0.01). Biodegradable PLLA sirolimus-eluting stents significantly reduced neointimal area at 180 days compared to permanent polymer Cypher stents (1.8 vs 3.0 mm²; p=0.01) in a porcine model.