Key result
IQ-SPECT underestimates true EDV by ~41%, yielding systematically smaller volumes than conventional LEHR.
Why the study?
Does IQ-SPECT compared to LEHR accurately estimate ejection fraction and left ventricular volumes in a dynamic heart phantom?
Does IQ-SPECT compared to LEHR accurately estimate ejection fraction and left ventricular volumes in a dynamic heart phantom?
IQ-SPECT produces systematically smaller left ventricular volumes than conventional LEHR MPI protocols, indicating that volume estimates are highly dependent on reconstruction parameters and software.
IQ-SPECT EF and LV volume results may differ from LEHR; leaves open need for protocol-specific validation before routine MPI use.
Developments in single photon emission tomography instrumentation and reconstruction methods present a potential for decreasing acquisition times. One of such recent options for myocardial perfusion imaging (MPI) is IQ-SPECT. This study was motivated by the inconsistency in the reported ejection fraction (EF) and left ventricular (LV) volume results between IQ-SPECT and more conventional low-energy high-resolution (LEHR) collimation protocols. IQ-SPECT and LEHR quantitative results were compared while the equivalent number of iterations (EI) was varied. The end-diastolic (EDV) and end-systolic volumes (ESV) and the derived EF values were investigated. A dynamic heart phantom was used to produce repeatable ESVs, EDVs and EFs. Phantom performance was verified by comparing the set EF values to those measured from a gated multi-slice X-ray computed tomography (CT) scan (EF True ). The phantom with an EF setting of 45, 55, 65 and 70% was imaged with both IQ-SPECT and LEHR protocols. The data were reconstructed with different EI, and two commonly used clinical myocardium delineation software were used to evaluate the LV volumes. The CT verification showed that the phantom EF settings were repeatable and accurate with the EF True being within 1% point from the manufacture’s nominal value. Depending on EI both MPI protocols can be made to produce correct EF estimates, but IQ-SPECT protocol produced on average 41 and 42% smaller EDV and ESV when compared to the phantom’s volumes, while LEHR protocol underestimated volumes by 24 and 21%, respectively. The volume results were largely similar between the delineation methods used. The reconstruction parameters can greatly affect the volume estimates obtained from perfusion studies. IQ-SPECT produces systematically smaller LV volumes than the conventional LEHR MPI protocol. The volume estimates are also software dependent.
No takes yet. Share an insight, caveat, or question.
Hippeläinen et al. (2017) studied Myocardial perfusion imaging (phantom study). IQ-SPECT protocol vs. LEHR protocol was evaluated on Underestimation of end-diastolic volume (EDV) and end-systolic volume (ESV) compared to true phantom volumes. IQ-SPECT produced systematically smaller left ventricular volumes, underestimating EDV by 41% and ESV by 42%, compared to 24% and 21% underestimation by the conventional LEHR protocol.