PURPOSE: To design and optimize a minibeam collimator for minibeam radiation therapy studies using a 250 kVp x-ray machine as a simulated synchrotron source. METHODS: A Philips RT250 orthovoltage x-ray machine was modeled using the EGSnrc/BEAMnrc Monte Carlo software. The resulting machine model was coupled to a model of a minibeam collimator with a beam aperture of 1 mm. Interaperture spacing and collimator thickness were varied to produce a minibeam with the desired peak-to-valley ratio. RESULTS: Proper design of a minibeam collimator with Monte Carlo methods requires detailed knowledge of the x-ray source setup. For a cathode-ray tube source, the beam spot size, target angle, and source shielding all determine the final valley-to-peak dose ratio. CONCLUSIONS: A minibeam collimator setup was created, which can deliver a 30 Gy peak dose minibeam radiation therapy treatment at depths less than 1 cm with a valley-to-peak dose ratio on the order of 23%.