A portable spectroelectrochemical sensor has been designed, evaluated, and demonstrated on a complex sample of radioactive waste. The sensor consisted of a black delrin sample compartment with a total internal sample volume of 800 microL, attached to an indium tin oxide coated glass multiple internal reflection optical element. Detection was by total internal reflection of light from a blue light emitting diode source. After a 10 min uptake for each standard, the sensor showed a linear response in absorbance change for 5 x 10(-5) to 5 x 10(-3) M ferrocyanide with electrochemical modulation by scanning at 20 mV/s from -0.30 V to +0.55 V vs a Ag/AgCl reference electrode. Due to the complex nature of Hanford radioactive tank waste samples containing ferrocyanide, a standard addition method was developed for analysis. The spectroelectrochemical sensor determined a concentration of 9.2 mM ferrocyanide for U-Plant-2 simulant solution containing 9.38 mM ferrocyanide that was prepared according to Hanford process flowsheets. A radioactive tank waste sample from Hanford Tank 241-C-112 was determined to be 1.0 mM in ferrocyanide using the spectroelectrochemical sensor. A value for the ferrocyanide concentration in the sample of 0.61 mM was determined by FTIR spectroscopy.