Coagulation upon blood-contacting biomaterials remains a problem for short- and long-term clinical applications. This study examined the ability of copper(II)-doped silicone surfaces to generate nitric oxide (NO) and locally inhibit coagulation. Silicone was doped with 3-μm copper [Cu(0)] particles yielding 3 to 10 weight percent (wt%) Cu in 70-μm thick Cu/silicone polymeric matrix composites (Cu/Si PMCs). At 3, 5, 8, and 10 wt% Cu doping, the surface expression of Cu was 12.1% ± 2.8%, 19.7% ± 5.4%, 29.0% ± 3.8%, and 33.8% ± 6.5%, respectively. After oxidizing Cu(0) to Cu(II) by spontaneous corrosion, NO flux, J(NO) (mol · cm(-2) · min(-1)), as measured by chemiluminescence, increased with surface Cu expression according to the relationship J(NO) = (1.63%SA(Cu) - 0.81) × 10(-11), R(2) = 0.98, where %SA(Cu) is the percentage of surface occupied by Cu. NO flux at 10 wt% Cu was 5.35 ± 0.74 × 10(-10) mol · cm(-2) · min(-1). The clotting time of sheep blood exposed to these surfaces was 80 ± 13 seconds with pure silicone and 339 ± 44 seconds when 10 wt% Cu(II) was added. Scanning electron microscopies (SEMs) of coatings showed clots occurred away from exposed Cu dendrites. In conclusion, Cu/Si PMCs inhibit coagulation in a dose-dependent fashion related to the extent of copper exposure on the coated surface.
No takes yet. Share an insight, caveat, or question.
Amoako et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: