Key result
Dual-chamber pacemaker implantation via the cephalic vein proves feasible in equids despite atrial lead dislodgment.
Why the study?
Is dual-chamber pacemaker implantation via the cephalic vein feasible and safe in healthy equids?
Is dual-chamber pacemaker implantation via the cephalic vein feasible and safe in healthy equids?
Dual-chamber pacemaker implantation via the cephalic vein is a feasible technique in equids, though atrial lead dislodgment remains a risk.
Answers are grounded in this paper and cite their sources.
Provides proof-of-concept in equids; hypothesis-generating and should not change clinical practice.
The purpose of the present study was to develop a feasible and safe technique for dual-chamber pacemaker implantation in healthy horses. Implantation was performed in a standing, tranquilized horse and in ponies. Atrial and ventricular leads were transvenously inserted through the cephalic vein, and a subcutaneous pacemaker pocket was created between the lateral pectoral groove and the manubrium sterni in 6 equids. Positioning of each lead was guided by echocardiography and by measuring the electrical characteristics of the lead. The implantation procedure lasted about 4 hours in each animal and was well tolerated. In all animals, dual-chamber pacemaker function was obtained, and these results remained good throughout the follow-up period. At the time of implantation, atrial and ventricular sensing were between 2.1 and 7.2 mV and 7.8 and 16.8 mV, respectively, and atrial and ventricular pacing thresholds at 0.5 millisecond varied from 0.5 to 0.7 V and from 0.3 to 1.0 V, respectively. Six months after the implantation, sensing values varied from 2 to 10 mV for the atrial lead and from 2 to 16 mV for the ventricular lead, while pacing thresholds at 0.5 millisecond varied from less than 0.5 to 2.5 V for the right atrium and from less than 0.5 to 5.0 V for the right ventricle. Atrial lead dislodgment occurred in 2 animals, requiring insertion of a new lead. Ventricular lead dislodgment was not observed.
No takes yet. Share an insight, caveat, or question.
Loon et al. (2001) studied Healthy equids (n=6). Dual-chamber pacemaker implantation via the cephalic vein was evaluated on Feasibility and safety (pacemaker function and lead stability). Dual-chamber pacemaker implantation via the cephalic vein was feasible in 6 healthy equids, maintaining good function over 6 months, although atrial lead dislodgment occurred in 2 animals.