New modes of nanoelectrospray (nES) have been identified that exhibit stable high-frequency current oscillations in the kilohertz range. In-line flow rate measurements during nES have for the first time allowed the accurate determination of the spray current and volumetric flow rate dependence on applied voltage. High-speed video imaging has revealed that each current pulse corresponds to the periodic formation of a short-lived stable cone-jet structure at the capillary tip. These findings are of particular importance and suggest greater process monitoring and control in nES-MS systems is necessary if efficiency and ion yield are to be optimized.
No takes yet. Share an insight, caveat, or question.
Alexander et al. (2006) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: