The authors evaluated the analytic and clinical performance of a sensitive radial partition fluorescent enzyme immunoassay for thyrotropin (TSH) performed on Stratus and compared it with a nonsensitive radioimmunoassay (RIA) method. Sensitivity of 0.15 mIU/L was obtained, and precision, specificity, and linearity were acceptable. A good correlation was observed between the two assays in samples from 311 hospitalized patients (r = 0.976). Stratus TSH results were outside the reference range for 20% of clinically euthyroid patients (n = 126), and 2.4% had undetectable levels. The clinically hyperthyroid group (n = 11) with the exception of one patient had TSH values below 0.2 mIU/L. Only 39% of hypothyroid patients on thyroid hormone replacement (n = 74) had TSH values in the reference range, with 38% and 23% exhibiting low and high values, respectively. All untreated primary hypothyroid patients (n = 8) had elevated TSH concentrations. The authors conclude that this sensitive TSH assay is useful for diagnosing hyperthyroidism when there is a clinical suspicion but cannot be recommended for thyroid screening in hospitalized patients.
No takes yet. Share an insight, caveat, or question.
Sacks et al. (1990) studied this question.