During the last decade, there has been a great advance in the kidney dialysis system by wearable artificial kidney (WAK) system for end-stage renal disease patients. Uremic solute removal and water regeneration system are the most prerequisite for WAK to work properly. In this study, we designed a filtering membrane system by using immobilized urease silk fibroin filter and evaluated its comparative effectiveness with a PVDF filtering system in peritoneal dialysate regeneration system by urea removal efficacy. We evaluated this membrane's characteristic and performances by conducting SEM-EDX analyze, water-binding abilities and porosity test, removal abilities of urea, cytotoxicity assay and enzyme activity assay. Under the condition for optimization of urease, the percentage removal of urea was about 40% and 60% in 50 mg/dL urea solution by urease immobilized PVDF and silk fibroin scaffolds, respectively. The batch experimental result showed that immobilized filter removed more than 50% of urea in 50 mg/dL urea solution. In addition silk fibroin with urease filter removed 90 percent of urea in the peritoneal dialysate after 24 h filtration. We suggest that silk fibroin with urease fixation filter can be used more effectively for peritoneal dialysate regeneration system, which have hydrophilic property and prolonged enzyme activity.