Key result
Preload-responsive pulsatile pump maintains cardiac output ~1.5 L/min higher than centrifugal pump in canine VF.
Why the study?
Does a preload responsive pulsatile pump improve hemodynamics compared to a steady flow centrifugal pump in canines during ventricular fibrillation?
Does a preload responsive pulsatile pump improve hemodynamics compared to a steady flow centrifugal pump in canines during ventricular fibrillation?
Absolute Event Rate: 2.4% vs 0.9%
A preload responsive pulsatile pump maintains cardiac output and prevents systemic vascular resistance elevation better than a steady flow centrifugal pump during total cardiac support in a canine model.
Preclinical data suggest hemodynamic benefit in VF arrest; leaves open translation to human mechanical support.
A unique preload responsive pulsatile pump was compared to a centrifugal pump in total cardiac support in 25-kg canines (n = 6, each group) in the left atrial-to-aorta mode during 5 h of ventricular fibrillation. With steady flow, there was immediate drop in output from 2.1 +/- 1.0 L/min to 1.4 +/- 0.3 L/min, followed by further reduction to 0.9 +/- 0.2 L/min during 5 h of ventricular fibrillation. With a pulsatile pump, there was no significant reduction from control of 2.4 +/- 0.6 L/min and no decline during 5 h of ventricular fibrillation. With steady flow, systemic vascular resistance (SVR) rose significantly from 1,762 dyne-s-cm-5 immediately on pump to 3,013 dyne-s-cm-5 at 5 h. With physiologic pulsatile flow, significant elevation of SVR did not occur. When stressed, due to diminished left atrial return, the centrifugal pump displayed line chatter and streaks of microbubbles, whereas the pulsatile pump did not. Crystalloid volume replacement with the centrifugal pump was 6.5 +/- 1.9 L, and with the preload responsive pulsatile pump, 5.6 +/- 1.3 L. It is concluded that in the left atrial-to-aorta mode during 5 h of ventricular fibrillation and with comparable volume replacement, total cardiac support of canines is associated with lower SVR with physiologic pulsatile flow and is not accompanied by line chatter and cavitation with this preload responsive pump.
No takes yet. Share an insight, caveat, or question.
Runge et al. (1991) studied Ventricular fibrillation (n=12). Preload responsive pulsatile pump vs. Centrifugal pump (steady flow) was evaluated on Cardiac output at 5 hours (L/min). In a canine model of ventricular fibrillation, a preload responsive pulsatile pump maintained cardiac output at 2.4 L/min compared to a drop to 0.9 L/min with a steady flow centrifugal pump.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: