A method to simultaneously determinate 19 chlorophenols (CPs) in water by liquid chromatography, coupled with atmospheric pressure chemical ionization mass spectrometry (LC‐APCI‐MS), and a solid‐phase extraction utilizing Oasis® HLB cartridges to preconcentrate the target analytes was developed. The method was linear in the range 0.5∼25.0 µg/L with a good correlation coefficient (r>0.996) for all CPs. The limits of detection (LODs) were 0.4∼30.0 ng/L, extraction recoveries were 87.5∼100.4%, intra‐day RSDs were 5∼12%, and inter‐day RSDs were 7∼15%. This method has been successfully applied to the analysis of water samples. In addition, a comparison was obtained between APCI‐MS and ESI‐MS for the determination of the 19 CPs.
No takes yet. Share an insight, caveat, or question.
Jin et al. (2006) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: