An experimental investigation has been carried out to determine the extent to which axial divergence in a conventional powder diffractometer influences the measurement of peak profile parameters. A Siemens D5000 θ–2θ diffractometer was used for this study along with the LaB 6 line profile standard SRM 660. Eight unique levels of axial divergence were investigated either by removing one or more of the Soller slits, or by introducing different combinations of Soller slits in the incident and diffracted beams. The measure of axial divergence used throughout is based on the maximum axial divergences of the incident and diffracted beams, Ψ i and Ψ d . Axial divergence produces a small and almost constant shift δ2θ p in the peak angle which in a typical diffractometer would amount to a zero offset 2θ≈0.005°. The integrated intensity of a profile increases almost linearly with the product Ψ i * Ψ d . The increase in breadth in profiles in the angular region 2θ<120° arises mainly from the change in asymmetry D H = H lo − H hi in which one side of the profile broadens (i.e., either the high angle or low angle side) without any significant broadening of the other side. Moreover, the asymmetry D H is linearly dependent on 2θ for a fixed level of axial divergence, and linearly dependent on the total axial divergence Ψ i +Ψ d at fixed 2θ.