Answers are grounded in this paper and cite their sources.
We describe a general method to determine the location of a point source of waves relative to a two-dimensional single-crystalline active pixel detector. Based on the inherent structural sensitivity of crystalline sensor materials, characteristic detector diffraction patterns can be used to triangulate the location of a wave emitter. The principle described here can be applied to various types of waves, provided that the detector elements are suitably structured. As a prototypical practical application of the general detection principle, a digital hybrid pixel detector is used to localize a source of electrons for Kikuchi diffraction pattern measurements in the scanning electron microscope. This approach provides a promising alternative method to calibrate Kikuchi patterns for accurate measurements of microstructural crystal orientations, strains, and phase distributions.
No takes yet. Share an insight, caveat, or question.
Vespucci et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: