Answers are grounded in this paper and cite their sources.
Piezoelectric biosensors have the potential to provide direct detection of food contaminants, such as pathogens. In this study, Protein A antibody immobilization was used for the activation of the piezoelectric biosensor to detect Salmonella typhimurium. The overall system consisted of a new design for a flow cell and flow injection analysis system. The flow cell made possible a baseline stability of ± 1 Hz out of 5 MHz for hours. The sensor had responses of 5 to 65 Hz in 30 min with R 2 = 0.95 for S. typhimurium concentrations of 10 7 to 10 9 CFU/ml under continuous flow, and 3 to 75 Hz in 40 min with R 2 = 0.96 for S. typhimurium concentrations of 10 6 to 10 10 CFU/ ml under stop flow. Cross‐reactivity tests were performed with nonpathogenic Escherichia coli, Escherichia coli O157:H7, Listeria monocytogenes , and Vibrio parahaemolyticus and showed less than 10% response.
No takes yet. Share an insight, caveat, or question.
Babacan et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: