Answers are grounded in this paper and cite their sources.
V-block refractometers have been used since the 1940s to retrieve the refractive index values of substances or optical glasses. When used outside laboratories, they are submitted to temperature variations which degrade their accuracy by varying the refractive index of the glasses and the length of the prisms. This paper proposes a method to calibrate a double-prism V-block refractometer by retrieving the values of two coefficients at a constant temperature and by applying corrections to these coefficients when the instrument is used at different temperatures. This method is applied to calibrate in salinity a NOSS instrument which can be used at sea on drifting floats, and the results show that measurement errors can be reduced by a factor of 5.8.
No takes yet. Share an insight, caveat, or question.
Marc Le Menn (2017) studied this question.