For part I see El Harrous et. al., ibid., vol.5, pp.1065-70 (1994). This paper continues the description of an isothermal titration calorimeter already outlined in the previous paper. We have determined its thermal parameters and report here on the performance of the calorimeter and the method of measuring via electrical compensation of the thermal effect. We have also calculated the effective volume of the reaction vessel and the heat of stirring of the reaction mixture in the vessel, and of injecting water into water. Finally, correct working of the calorimeter has been tested by a dilution experiment of a saccharose solution.
No takes yet. Share an insight, caveat, or question.
Harrous et al. (1994) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: