Answers are grounded in this paper and cite their sources.
Urine specimens from pain management patients dosed with Nucynta (Tapentadol) were confirmed for the presence of tapentadol and N-desmethyltapentadol using ultra-performance liquid chromatography-tandem mass spectrometry to minimize sample preparation and urine volume requirements. The linearity of the method for both tapentadol and N-desmethyltapentadol demonstrated correlation coefficients (R²) above 0.99 and linear ranges from 50 to 500,000 ng/mL for tapentadol and 100 to 500,000 ng/mL for N-desmethyltapentadol. The intraday precision of the assay for both analytes ranged from 2.2 to 6.9% over three concentrations; the interday precision for both analytes ranged from 1.2 to 8.4%. The limits of quantitation were 50 and 100 ng/mL for tapentadol and N-desmethyltapentadol, respectively, and the upper limit of linearity for both analytes was determined to be 500,000 ng/mL. Urine samples were collected within 24 h of dosing with tapentadol and shipped overnight to the laboratory. Samples were hydrolyzed with acid prior to analysis to measure total (unconjugated and conjugated) tapentadol and N-desmethyltapentadol. Further investigation into characterization of metabolites was performed by using a hybrid quadrupole-time-of-flight mass spectrometer in lieu of suitable analytical reference standards. The presence of significant N-desmethyltapentadol glucuronide was demonstrated for the first time.
No takes yet. Share an insight, caveat, or question.
Bourland et al. (2010) studied this question.