An inverse methodology for the design of biologically loaded radio-frequency (RF) coils for magnetic resonance imaging applications is described. Free space time-harmonic electromagnetic Green's functions and de-emphasized B1 target fields are used to calculate the current density on the coil cylinder. In theory, with the B1 field de-emphasized in the middle of the RF transverse plane, the calculated current distribution can generate an internal magnetic field that can reduce the central overemphasis effect caused by field/tissue interactions at high frequencies. The current distribution of a head coil operating at 4 T (170 MHz) is calculated using an inverse methodology with de-emphasized B1 target fields. An in-house finite-difference time-domain routine is employed to evaluate B1 field and signal intensity inside a homogenous cylindrical phantom and then a complete human head model. A comparison with a conventional RF birdcage coil is carried out and demonstrates that this method can help in decreasing the normal bright region caused by field/tissue interactions in head images at 170 MHz and higher field strengths.
No takes yet. Share an insight, caveat, or question.
Bin et al. (2005) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: