Key result
HCL system maintains optimal perioperative glycemia in a pediatric T1D case without requiring IV insulin.
Why the study?
The latest guidelines for managing children with T1D undergoing surgery were published before hybrid closed-loop systems became common, leaving uncertainty about their perioperative use.
Does continuing a hybrid closed-loop insulin system maintain optimal perioperative glycemia in a pediatric patient with type 1 diabetes undergoing surgery?
Case Report (n=1)
No
Does continuing a hybrid closed-loop insulin system maintain optimal perioperative glycemia in a pediatric patient with type 1 diabetes undergoing surgery?
Continuing a hybrid closed-loop insulin system during major surgery in a pediatric patient with type 1 diabetes safely maintained optimal glycemic control without requiring intravenous insulin or hourly monitoring.
The goal in type 1 diabetes (T1D) therapy is to maintain optimal glycemic control under any circumstance. Diabetes technology is in continuous development to achieve this goal. The most advanced Food and Drug Administration- and European Medicines Agency-approved devices are hybrid closed-loop (HCL) systems, which deliver insulin subcutaneously in response to glucose levels according to an automated algorithm. T1D is frequently encountered in the perioperative period. The latest international guidelines for the management of children with diabetes undergoing surgery include specific adjustments to the patient's insulin therapy, hourly blood glucose monitoring, and intravenous (IV) insulin infusion. However, these guidelines were published while the HCL systems were still marginal. We present a case of a 9-year-old boy with long-standing T1D, under HCL system therapy for the last 9 months, and needing surgery for an appendectomy. We agreed with the family, the surgical team, and the anesthesiologists to continue HCL insulin infusion, without further adjustments, hourly blood glucose checks or IV insulin, while monitoring closely. The HCL system was able to keep glycemia within range for the total duration of the overnight fast, the surgery, and the initial recovery, without any external intervention or correction bolus. This is, to the best of our knowledge, the first reported pediatric case to undergo major surgery using a HCL system, and the results were absolutely satisfactory for the patient, his family, and the medical team. We believe that technology is ripe enough to advocate for a "take your pump to surgery" message, minimizing the impact and our interventions. The medical team may discuss this possibility with the family and patients.
No takes yet. Share an insight, caveat, or question.
Domínguez‐Riscart et al. (2021) conducted a case report in Type 1 Diabetes (n=1). Hybrid closed-loop system (Medtronic 670G) was evaluated on Perioperative glycemic control. The hybrid closed-loop system successfully maintained optimal perioperative glycemia (80-180 mg/dl) in a 9-year-old boy with type 1 diabetes undergoing an appendectomy without requiring external adjustments or intravenous insulin.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: