Calibration study reveals greater than 99.99% detection efficiency for charged particles in the DAMPE satellite, highlighting high-precision on-orbit tracking and cosmic-ray discrimination.
The DArk Matter Particle Explorer ( DAMPE ) is a space-borne apparatus for detecting the high-energy cosmic-ray-like electrons, γ -rays, protons and heavy ions. The Plastic Scintillator Detector (PSD) is the top-most sub-detector of the DAMPE . The PSD is designed to measure the charge of incident high-energy particles and it also serves as a veto detector for discriminating γ -rays from charged particles. In this paper, a PSD on-orbit calibration procedure is described, which includes the five steps of pedestal, dynode correlation, response to minimum-ionizing particles, light attenuation function and energy reconstruction. A method for reconstructing the charge of incident high energy cosmic-ray particles is introduced. The detection efficiency of each PSD strip is verified to be above 99.5%; the total efficiency of the PSD for charged particles is above 99.99%.