A novel, microporous, hydrophilic polyether urethane-urea (PEUU) vascular graft was compared with expanded PTFE in the canine carotid artery. At implantation times ranging from 4 h to 6 months, all the PEUU grafts were found to be occluded while of the PTFE grafts, only those implanted for 1 week and 6 months were blocked. Histopathological analysis of the explanted grafts and their capsules revealed an ongoing inflammatory reaction at the anastomotic sites of the PEUU grafts.
No takes yet. Share an insight, caveat, or question.
Martz et al. (1988) studied this question.