= 28). Specificity was evaluated for 42 patients with previous viral and parasitic diseases as well as a high level of rheumatic factor. The sensitivity of OG was 95.8% (95% confidence interval [CI95%], 89.6 to 98.8) for samples collected ≥10 days after the onset of symptoms, which was equivalent to the sensitivity of ASIA of 90.5% (CI95%, 82.8 to 95.6). OG uncovered six false-negative results of ASIA, of which two had only IgM with OG. Specificity was 100% (CI95%, 93.4 to 100) with both tests on samples, including patients infected with endemic coronavirus. Overall, OG performance characteristics indicate that the test is suitable for routine use in clinical laboratories, and its performance is equivalent to that of immunoassay. Testing OG on a larger asymptomatic population may be needed to confirm these results.