An RP-HPLC method for the analysis of adenosine (ADO) has been developed and validated. In the present study, we report an RP-HPLC-based method with modifications of mobile phase and shorter retention time that substantially improved the efficiency of ADO analysis. The HPLC separation of the ADO was achieved on a C18 column, using a mobile phase consisting of water, containing 7% v/v ACN, at a flow rate of 0.8 mL/min. The column effluent was monitored by UV detection at 260 nm. A linear response was achieved over the concentration range of 0.25-100.00 micromol/L. The analytical method inter- and intra-run accuracy and precision were better than +/- 15%. The LOQ was 0.25 micromol/L, with ADO detection in the range of 6.25 pmol per sample. The method has been applied to the study of adenosine kinase (AK) kinetics.
No takes yet. Share an insight, caveat, or question.
Marin et al. (2007) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: