A fully inkjet-printed disposable and low cost paper-based device for potentiometric Na + - or K + -ion sensing has been developed. A printed ionophore-based all-solid-state ion selective electrode on a graphene/poly(3,4-ethylenedioxythiophene) polystyrenesulfonate (G/PEDOT:PSS) nanocomposite solid contact and a printed all-solid state reference electrode consisting of a pseudosilver/silver chloride electrode coated by a lipophilic salt-incorporating poly(vinyl chloride) membrane overprinted with potassium chloride have been combined on a microfluidically patterned paper substrate. Devices are built on standard filter paper using off-the-shelf materials. Ion sensing has been achieved within 180 s by simple addition of 20 μL of sample solution without electrode preconditioning. The limits of detection were 32 and 101 μM for Na + and K +, respectively. The individual single-use sensing devices showed near Nernstian response of 62.5 ± 2.1 mV/decade (Na + ) and 62.9 ± 1.1 mV/decade (K + ) with excellent standard potential ( E 0 ) reproducibilities of 455.7 ± 5.1 mV (Na + ) and 433.9 ± 2.8 mV (K + ). The current work demonstrates the promising possibility of obtaining low-cost and disposable paper-based potentiometric sensing devices potentially manufacturable at large scales with industrial inkjet printing technology.
No takes yet. Share an insight, caveat, or question.
Ruecha et al. (2017) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: