Reproducibility in mass spectral data is important in both biomarker discovery and spectral database searching. We report a strategy, employing a series of substituted benzylpyridinium thermometer ions that can be used to monitor changes in performance of multiple aspects of an electrospray ionization source that impact the intensity axis of a spectrum. Performance attributes, which could confound even isotope-based quantification strategies, are readily assessed using a mixture of thermometer ions. Based on the observed behavior of the ions, a procedure is proposed for monitoring instrument performance and compensating for factors that affect reproducibility across both time and instruments.
No takes yet. Share an insight, caveat, or question.
Lecchi et al. (2008) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: