Key result
Simulated smaller pacemaker helix wires worsen early myocardial micro-damage, while larger helices increase overall damage.
Why the study?
The risk of myocardial penetration from active-fixation screw-in pacing leads increases as helix electrodes become smaller, but contributing factors for lead penetration are not fully understood.
Finite element analysis demonstrates that pacemaker lead helix wire diameter and helix diameter significantly affect the timing and extent of acute myocardial micro-damage.
Answers are grounded in this paper and cite their sources.
Helix dimensions may inform lead design to limit micro-damage; leaves open in vivo validation and clinical translation.
The risk of myocardial penetration due to active-fixation screw-in type pacing leads has been reported to increase as the helix electrodes become smaller. In order to understand the contributing factors for lead penetration, we conducted finite element analyses of acute myocardial micro-damage induced by a pacemaker lead screw-in helix electrode. We compared the propensity for myocardial micro-damage of seven lead designs including a baseline model, three modified designs with various helix wire cross-sectional diameters, and three modified designs with different helix diameters. The comparisons show that electrodes with a smaller helix wire diameter cause more severe micro-damage to the myocardium in the early stage. The damage severity, represented by the volume of failed elements, is roughly the same in the middle stage, whereas in the later stage the larger helix wire diameter generally causes more severe damage. The onset of myocardial damage is not significantly affected by the helix diameter. As the helix diameter increases, however, the extent of myocardial damage increases accordingly. The present findings identified several of the major risk factors for myocardial damage whose consideration for lead use and design might improve acute and chronic lead performance.
No takes yet. Share an insight, caveat, or question.
Zhao et al. (2011) studied Myocardial micro-damage. Modified pacemaker lead designs with various helix wire and helix diameters vs. Baseline model was evaluated on Myocardial micro-damage (volume of failed elements). Pacemaker lead electrodes with smaller helix wire diameters caused more severe early myocardial micro-damage, while larger helix diameters increased the overall extent of damage in simulation models.
Synapse has enriched 10 related papers. Consider them for comparative context: