This brief presents a low-power analog acquisition front-end circuit for a Wireless Body Area Network. This front-end system mainly consists of three parts, namely, chopped capacitively coupled instrumentation amplifier (CCIA), switched capacitor filter (SC-filter), and successive-approximation analog-to-digital converter. In order to reduce the power consumption, the supply voltage is scaled to 0.5 V, and all analog building blocks are biased in the subthreshold region. The chopper-stabilized technique is introduced to eliminate the 1/f noise, and a dc-servo loop is employed in the CCIA to suppress the electrode offset. A low-power second-order SC-filter is employed to eliminate the spikes produced by the CCIA, which also realizes a tunable gain to satisfy the specification. This low-power analog front-end circuit has been fabricated in a 0.18-μW CMOS process. It occupies 1 mm2and consumes a minimal 1.3 μW at 0.5 V. It achieves a bandwidth of 0.5-250 Hz, a CMRR of 95 dB, and an input impedance of 48 MΩ, respectively.
No takes yet. Share an insight, caveat, or question.
Zhu et al. (2016) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: