Key result
Titan Optimax biostent shows similar 1-year major event rates compared to the Titan-2 stent.
Why the study?
The safety and efficacy of the new cobalt-chromium bioactive stent Titan Optimax compared with its predecessor Titan-2 in real-world PCI patients were not established.
Does the Titan Optimax stent reduce the composite of death, AMI, TLR, or stent thrombosis compared to the Titan-2 stent in patients undergoing percutaneous coronary intervention?
Cohort (n=784)
Yes
Does the Titan Optimax stent reduce the composite of death, AMI, TLR, or stent thrombosis compared to the Titan-2 stent in patients undergoing percutaneous coronary intervention?
Absolute Event Rate: 7.6% vs 10.3%
p-value: p=0.211
The Titan Optimax stent demonstrates comparable safety and efficacy to its predecessor Titan-2 at 1 year, with potentially improved outcomes in reducing non-fatal events among patients with STEACS.
Comparable 1-year outcomes support Optimax safety versus Titan-2 in real-world PCI; hypothesis-generating for STEACS benefit.
Objectives To compare the safety and efficacy of the new cobalt‐chromium bioactive stent Titan Optimax® (Hexacath, France) with its predecessor, Titan‐2®. Background The TIOMAX registry includes 784 patients who underwent percutaneous coronary intervention with these stents in 21 Spanish hospitals. Methods Analysis of all patients in the registry without exclusion criteria, candidates for revascularization (March‐2013/July‐2014). Initially 273 patients received Titan‐2®, and the next 511 received the Optimax® after its launch. Results Mean age was 65.8 ± 13.0 (78.1% men); 49.2% were STEACS patients (n = 322), 29.8% NSTEACS, and 27.3% had stable angina or silent ischemia. Most STEACS patients (76.4% of n = 322) were treated <24 hr after developing symptoms. All‐cause death (D), cardiac death (CD), acute myocardial infarction (AMI), and stent thrombosis (ST) at 1 month were 1.1, 0.8, 0.1, and 0.5%, respectively, with no significant differences between groups. At 1 year, the death rate was 5.5% for Titan‐2 vs. 4.1% for Optimax®, CD was 1.8% for both groups, ST 1.1 vs. 0.6%, new AMI 3.3 vs. 2.5% and target lesion revascularization (TLR) 3.7 vs. 2.9%. The primary endpoint of the composite event (CE) of D/AMI/TLR/ST occurred in 10.3% vs. 7.6% (p = 0.211). Patients with STEACS (N = 322: Titan‐2/Optimax: 103/209) had better outcomes for secondary events, device‐oriented failure CD/AMI/TLR (7.8% vs. 5.0%; p = 0.330), and non‐fatal CE of AMI/ST/TLR (7.8% vs. 2.7%, p = 0.039). Conclusions The Titan Optimax retains the efficacy and safety of Titan 2. It appears to perform better in the subgroup of STEACS patients, by reducing the non‐fatal CE of AMI/ST/TLR.
No takes yet. Share an insight, caveat, or question.
López-Mínguez et al. (2017) conducted a cohort in Coronary artery disease requiring percutaneous coronary intervention (n=784). Titan Optimax biostent vs. Titan-2 biostent was evaluated on Composite event of death, acute myocardial infarction, target lesion revascularization, and stent thrombosis at 1 year (p=0.211). In a real-world registry, the Titan Optimax biostent had similar 1-year rates of death, AMI, TLR, or stent thrombosis compared to the Titan-2 stent (7.6% vs 10.3%; P=0.211).