It has been suggested that electrode dissolution and the concomitant saturation of the electrode's diffusion layer restricts the detection limit of the jalpaite Cu ion-selective electrode (ISE) to samples with total Cu levels above 10(exp-5) mol dm(exp -3) [1, 2]. This article will use rotating disk electrode (RDE) data for San Diego Bay seawater and Fick's law of diffusion to demonstrate that the static commercial Orion Cu ISE (employing a jalpaite membrane) produces a background level of contamination of (2.0 plus or minus 0.5) x 10(exp -8) mol dm(exp -3) total Cu, and the reduced thickness of the Orion Cu ISE's diffusion layer in the presence of hydrodynamic flow [e.g., at an RDE, or in continuous flow analysis (CFA)] lowers the background contamination of Cu to < 10(exp-9) mol dm(exp -3). Furthermore, the RDE Cu ISE employing an electrode fabricated using jalpaite precipitated in 80% excess Na2S, so as to minimize the presence of occluded and leachable Cu salts and extraneous phases such as silver sulfide [3 - 5], reveals an improvement in the lower limit of detection compared to the commercial Orion Cu ISE.
No takes yet. Share an insight, caveat, or question.
Zirino et al. (2002) studied this question.