Key result
New high-power time-domain fNIRS device achieves reliable 20 Hz acquisition to detect brain resting-state oscillations.
Why the study?
Previous analogue TD fNIRS devices had lower signal-to-noise ratio and limited acquisition rates, motivating development of a high power setup with improved performance.
A newly developed TD fNIRS device allows for fast 20 Hz acquisition rates and improved signal-to-noise ratio for brain resting-state oscillation studies.
Answers are grounded in this paper and cite their sources.
Advances fNIRS temporal resolution for research; should not yet change clinical practice.
A high power setup for multichannel time-domain (TD) functional near infrared spectroscopy (fNIRS) measurements with high efficiency detection system was developed. It was fully characterized based on international performance assessment protocols for diffuse optics instruments, showing an improvement of the signal-to-noise ratio (SNR) with respect to previous analogue devices, and allowing acquisition of signals with sampling rate up to 20 Hz and source-detector distance up to 5 cm. A resting-state measurement on the motor cortex of a healthy volunteer was performed with an acquisition rate of 20 Hz at a 4 cm source-detector distance. The power spectrum for the cortical oxy- and deoxyhemoglobin is also provided.
No takes yet. Share an insight, caveat, or question.
Re et al. (2022) studied Healthy (n=1). High power time-domain (TD) fNIRS device was evaluated on Device performance (acquisition rate and signal-to-noise ratio). A newly developed high-power time-domain fNIRS device achieved a reliable 20 Hz acquisition rate at source-detector distances up to 5 cm, successfully detecting brain resting-state oscillations.