A reversed-phase high-performance liquid chromatographic procedure for the simultaneous determination of impurities associated with pseudoephedrine hydrochloride (PSEH) and naproxen sodium (NapNa) is developed and validated. The method is developed using a Waters Spherisorb cyano column (5 microm, 250 x 4.6 mm). An isocratic elution in a water-acetonitrile-methanol-triethylamine mixture (850:75:75:5) is adjusted to a pH of 3.7 +/- 0.02 with formic acid as the mobile phase. The UV detection was set at 260 nm, and the wavelength was switched to 235 nm before the elution of the last component, 2-ethyl-6-methoxy-naphthalene (EMN). The method is shown to be linear at a concentration range of 0.24 to 1.92 microg/mL for benzaldehyde, benzoic acid, and 2-(methylamino)-propiophenone hydrochloride, which are known impurities of PSEH. The NapNa impurities, 2-(6'-hydroxy-2'-naphthyl) propionic acid, 2-hydroxy-6-methoxy-naphthalene, 1-(6'-methoxy-2'-naphthyl) ethanol, 2-acetyl-6-methoxy-naphthalene, and EMN are also demonstrated to be linear at a concentration range of 0.44 to 3.52 microg/mL. Under the chromatographic conditions of the method, all impurities are resolved from the active components.
No takes yet. Share an insight, caveat, or question.
Ekpe et al. (2001) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: