We have developed a planar optofluidic chip for the analysis of tumor cell apoptosis by exciting dual fluorescence through the integrated fibers. In contrast to conventional cell apoptosis study that requires commercial flow cytometer with a very high price tag and bulky optics, our design was realized via a simple single-layer soft lithography fabrication process. The device performance was tested by characterizing the HeLa cell apoptosis induced by hydrogen peroxide. The performance of the proposed optofluidic chip was concurrently confirmed by fluorescent microscopy and benchmarked against a commercial flow cytometer. This study demonstrated the capability of the proposed optofluidic chip for cell apoptosis assay. The major advantages of this device include simple fabrication process, compact optical components, reliable performance, and low manufacturing cost, making it a promising platform for future mass-producible, inexpensive, and disposable on-chip investigation of biological samples.
No takes yet. Share an insight, caveat, or question.
Guo et al. (2014) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: