Key result
ResNet and SqueezeNet using Gramain angular difference fields classify BCG signal quality with ~88% accuracy.
Why the study?
The BCG signal is susceptible to noise from motion artifacts, causing errors in blood pressure estimation and lowering measurement performance.
Time-series imaging combined with CNN models can effectively classify arm ballistocardiogram signal quality with up to 87.5% accuracy, potentially improving non-invasive continuous blood pressure measurement.
This paper describes a signal quality classification method for arm ballistocardiogram (BCG), which has the potential for non-invasive and continuous blood pressure measurement. An advantage of the BCG signal for wearable devices is that it can easily be measured using accelerometers. However, the BCG signal is also susceptible to noise caused by motion artifacts. This distortion leads to errors in blood pressure estimation, thereby lowering the performance of blood pressure measurement based on BCG. In this study, to prevent such performance degradation, a binary classification model was created to distinguish between high-quality versus low-quality BCG signals. To estimate the most accurate model, four time-series imaging methods (recurrence plot, the Gramain angular summation field, the Gramain angular difference field, and the Markov transition field) were studied to convert the temporal BCG signal associated with each heartbeat into a 448 × 448 pixel image, and the image was classified using CNN models such as ResNet, SqueezeNet, DenseNet, and LeNet. A total of 9626 BCG beats were used for training, validation, and testing. The experimental results showed that the ResNet and SqueezeNet models with the Gramain angular difference field method achieved a binary classification accuracy of up to 87.5%.
No takes yet. Share an insight, caveat, or question.
Shin et al. (2023) studied BCG signal quality assessment. Time-series imaging methods with CNN models was evaluated on Binary classification accuracy for high-quality versus low-quality BCG signals. ResNet and SqueezeNet models combined with the Gramain angular difference field method achieved a binary classification accuracy of up to 87.5% for distinguishing BCG signal quality.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: