To be able to describe more precisely the behavior of a real-time 3-D ultrasound system with either a dense array or various sparse designs, experimental data from a 2-D fully connected array prototype with 50 x 50 elements have been collected. The data have been processed off line to form synthetic aperture 3-D volume images. Simulated and experimental results are compared and show good correlation. The performance of the best sparse designs, all thinned to more than 50%, offer performances comparable to a dense array.
No takes yet. Share an insight, caveat, or question.
Austeng et al. (2002) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: