Conditioning skeletal muscle ventricles at 2 Hz yields poor hydraulic performance due to slow contractility.
Why the study?
Little is known about the power losses and hydraulic performance of skeletal muscle ventricles formed from conditioned skeletal muscle acting on circulating blood.
Does electrical conditioning of sheep latissimus dorsi muscles improve the hydraulic performance of skeletal muscle ventricles?
Does electrical conditioning of sheep latissimus dorsi muscles improve the hydraulic performance of skeletal muscle ventricles?
Skeletal muscle ventricles constructed from electrically conditioned latissimus dorsi muscles show poor hydraulic performance due to slow contractile properties, indicating a need for refined stimulation protocols.
Answers are grounded in this paper and cite their sources.
Acute sheep muscle pump data warrant clinical caution; leaves open viability of conditioned skeletal muscle for cardiac assistance.
BACKGROUND: The fatigue resistance and power-to-weight ratio of skeletal muscle that has been conditioned by electrical stimulation makes cardiac assistance from a graft of such muscle a realistic prospect. A skeletal muscle must be surgically reconfigured to act on the circulating blood, but little is known about the power losses that accompany such interventions. We investigated in acute experiments the hydraulic performance of approximately cylindrical pumps made from sheep latissimus dorsi (LD) muscles, having first characterized the performance of each muscle in situ. METHODS AND RESULTS: Force-length and force-velocity relations were measured in situ for LD that had received either 8 weeks of stimulation at 2 Hz or no chronic stimulation. Two sizes of skeletal muscle ventricle (SMV) were formed from the same muscles, and their hydraulic performance was measured. The hydraulic performance was also calculated from the linear data, models of the force-length and force-velocity curves, and a description of the stress distribution within the SMV wall. The model predicted well the isovolumetric function of the ventricles and the optimum afterload but overestimated the flow and therefore the power. In conditioned ventricles the performance was particularly poor because of the slow contractile properties of the muscles. CONCLUSIONS: If SMVs are to pump effectively against the arterial impedance, the pressure drop caused by flow (or the internal resistance) should be lower than that of the ventricles we constructed. Progress can be made through refinement of surgical technique and stimulation protocols that generate faster fatigue-resistant muscles.
No takes yet. Share an insight, caveat, or question.
Jarvis et al. (1997) studied Cardiac assistance. Electrical stimulation vs. No chronic stimulation was evaluated on Hydraulic performance (flow and power). Skeletal muscle ventricles conditioned with 8 weeks of 2 Hz stimulation exhibited poor hydraulic performance due to the slow contractile properties of the muscles.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: