A newly designed, 100 mm2, silicon drift detector has been installed on an aberration-corrected scanning transmission electron microscope equipped with an ultra-high resolution pole piece, without requiring column modifications. With its unique, windowless design, the detector's active region is in close proximity to the sample, resulting in a dramatic increase in count rate, while demonstrating an increased sensitivity to low energy X-rays and a muted tilt dependence. Numerous examples of X-ray energy dispersive spectrometry are presented on relevant materials such as Al x Ga1-x N nanowires, perovskite oxides, and polycrystalline CdTe thin films, across both varying length scales and accelerating voltages.