Answers are grounded in this paper and cite their sources.
Recent critical applications for ultralow expansion materials have required thermal expansion measurements with very high precision and accuracy. A Fizeau interferometer employing a helium-neon laser has given a precision of +/-0.1 x 10(-6) cm/cm. A second, more rapid method consists of a rod type vitreous silica dilatometer. A stirred water bath is used for specimen temperature control and a high output, linear variable differential transformer serves as the extensometer. Precision of this method is +/-0.5 x 10(-6) cm/cm. Calibration procedures and results for several low expansion materials are discussed.
No takes yet. Share an insight, caveat, or question.
Plummer et al. (1968) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: