Laser-induced breakdown spectroscopy (LIBS) was evaluated to determine elements collected on swipes as surface contamination. A series of long laser plasmas formed along the swipe surface (Post-it paper) interrogated the collected contamination. LIBS detection limits, determined for the elements Ag, As, Ba, Be, Cd, Cr, Cu, Hg, Mn, Ni, Pb, Sr, and Zn on swipes (2 cm2 area), ranged from 0.002 μg (Be) to 1.46 μg (Pb). The elements were introduced as constituents of synthetic silicate particles serving as a contaminant dust stimulant. The average predicted mass was within 16% of the actual mass on the swipe. The efficiency of collecting particles from surfaces including plastic, Formica, and Al metal was also evaluated. The ability to detect and differentiate two amino acids on a swipe from each other and from the swipe using chemometric modeling techniques was also demonstrated.