Tapered optical fibers with nano-assembled coatings of thicknesses of order tens of nanometres were used for the detection of ammonia gas. The film coating was composed of alternate layers of tetrakis-(4-sulfophenyl) porphine (TSPP) and poly(allylamine hydrochloride) (PAH), which were deposited using the electrostatic self-assembly process (ESA). Exposure of a PAH/TSPP nano-assembled non-adiabatic tapered optical fiber with a waist diameter of 10 µm to ammonia induced significant optical changes in the transmission spectrum of the optical fiber. The fiber optic sensor showed a linear sensitivity to the concentration of ammonia in the range of 10–100 ppm, with response and recovery times less than 100 and 240 sec, respectively. The 3σ limit of detection (LOD) was estimated to be ca. 2 ppm.