Prospective multicenter single-arm study demonstrates technical success filling aneurysm sacs with shape memory polymer during endovascular repair, indicating viable sac embolization.
Preprocedural image analysis and intraprocedural techniques to fully treat infrarenal abdominal aortic aneurysm sacs outside of the endograft with shape memory polymer (SMP) devices during endovascular aneurysm repair were developed. Prospective, multicenter, single-arm studies were performed. SMP is a porous, self-expanding polyurethane polymer material. Target lumen volumes (aortic flow lumen volume minus endograft volume) were estimated from the preprocedural imaging studies and endograft dimensions. SMP was delivered immediately after endograft deployment via a 6F sheath jailed in a bowed position in the sac. Technical success was achieved in all cases, defined as implanting enough fully expanded SMP volume to treat the actual target lumen volume.
No takes yet. Share an insight, caveat, or question.
Holden et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: